General
CO2 Detection Method: Non-Dispersive Infrared Optical Sensor with patented gold plated optics and patented solid state source and detector.
Sample Method: Diffusion or Flow Through
Transmitter Rated Life: minimum 15 years
Operating Conditions: 32 to 122° F (0 to 50°C), 0 to 95% RH (Standard), -13 to 131°F (-25 to 55°C) (extended range) Storage Conditions: -22 to 158° F (-30 to 70° C)
Performance
CO2 Measurement Range: 0-5%, 0-30%, 0-60%, 0-100%
CO2 Accuracy: +/- 3 ppm+/- 3% of reading. Non Linearity: < 1% of full scale Calibration: Automatic calibration built in.
Response Time: 4 seconds to 2 minutes (user configurable via filter type and application. Reading updated twice a second.
Pressure Dependence: 0.13% of reading per mm Hg.
Operating Pressure Range: 950 to 1050 bar.
Minimum Run Time: 10 Seconds on duty-cycle operation
Optional Temperature & Humidity Sensor Note: Only available as a serial output Method: RH - Capacitive, Temp - Band-gap
Range: -13 to 131°F (-25 to 55°C), o to 95% RH
Resolution: 0.08°C, 0.08% RG
Temp Accuracy: ±1.8°F (1°C)@ Oto 55°C, ±3.6°F (±2°C)
full range,
RH Accuracy: ±3%RH@ 68°C to 131°F (20°C to 55°C), ±
5% full range
Power
Input: 3.25 to 15 voe
Peak Current: 33 mA
Average Current:< 1.5 mA
Power Consumption: 3.5 mW
Wiring Connection: 2 x 5 - 0.1" header
Application
*Medical - Anesthetic instruments, respirators, oxygen-enrichers
*Biotechnology - Oxygen incubators
*Food industry - Refrigeration, greenhouses
*Safety - Air conditioners, oxygen detectors, fire detectors
*Long life - 2.5~10 years in ambient air
*Virtually no influence from CO2, CO, H2S, NOx, H2
*Low cost
*Operates in normal ambient temperatures
*Stable output signal
*No external power supply required for sensor operation
*No warmup time is required
Notes:
1) When calibrated at both 0% and 100% of O2, accuracy in the range from 0-100 O2 shall be within ±1% of full scale O2 reading .
2) Sensors should be used under conditions where
the air exchange is greater than 200~300ml per minute in order to obtain the response speed as shown in the Specifications.
3) Life expectancy at 20˚C in normal air (1013hPa
/ 20.7% O2)is defined as the period until sensor output drops to 70% of its original value.
Additional Information:
Product Details:
| Minimum Order Quantity | 1 Piece |
| Operating Voltage | 12V |
| Brand | LuxMed |
| Output Current ( in mA) | 0 to 3,3V |
| Warranty | 1 year |
| Country of Origin | Made in India |
LUXMED Dual Beam NDIR Co2 Sensor
Ultra Low Power CO2 Sensor For Laboratory Applications
The LUXMED-Co2 is a CO2 module for OEM applications that represents a breakthrough in low power gas detection making it ideal for battery powered, energy harvesting or wireless applications with tight energy budgets requiring long operating life. This Infrared LED based sensor offers a number of distinctive features over conventional CO2 sensors.
Key features of the LUXMED-Co2 CO2 sensor include:
Ο Continuous power consumption of 3.5 mW. With reasonable duty cycling (e.g. 15 min) battery life can be extended to 5+ years.
Ο Measurement ranges of 0-2%, 0-20%, 0-60%, 0-100% (custom range possible with volume order).
Ο 10 second warm up time.
Ο Shock and vibration resistant.
Ο Built in auto calibration feature.
Ο Options for on-board temperature and humidity measurement and extended temperature operating range of -13 to 131ºF (-25 to 55ºC).
Ο Simple, closed contact initiation of ambient or zero calibration.
Ο Runs cool, there is no heat generated that can interfere with temperature or humidity measurements
Ο Flow through adaptor available.
Conventional CO2 sensors utilize a small incandescent bulb that generates a lot of heat and light energy in addition to the infrared energy used for the gas measurement. Instead of a light bulb, the LUXMED-Co2 uses a small light emitting diode (LED) that generates infrared energy only at the wavelength needed to measure CO2. This is the reason for the dramatically reduced energy use of the LUXMED-Co2.
LED sensors have always been recognized as idea for low power gas detection but their development has been a great technical challenge. The LUXMED Co2 developed by a Scotland based company called Gas Sensing Solutions is the very first infrared LED product on the market that is stable enough and economical enough to be sold into measurement and control products.
The photo to the right shows the inside of the sensor with the LED emitter and photo detector at the top of the sensor. The yellow line shows the path of the infrared energy inside the sensor. The LED only emits infrared energy at the wavelength that is absorbed by CO2. As a result any decrease in light getting to the detector relates to how much CO2 is in
the air. Under normal operation the LED flashes 2 times per second but can be made to flash up to 20 times per second.
Additional Information:
Product Details:
| Minimum Order Quantity | 1 Piece |
| Warranty | 2 year |
| Country of Origin | Made in India |
Additional Information:
Product Details:
| Minimum Order Quantity | 1 Piece |
| Brand | LuxMed |
| Material | Steel |
| Weight | 300 Gram |
| Country of Origin | Made in India |
Additional Information:
Product Details:
| Minimum Order Quantity | 1 Piece |
| Brand | LuxMed |
| Warranty | 2 Year |
| Country of Origin | Made in India |
Additional Information: